Technical Specification
Modes of ventilation
Assist/Control (A/C)
Synchronized Intermittent Mandatory Ventilation (SIMV)
Spontaneous ventilation (SPONT)
Manual ventilation (Manual)
Combined: VCV + SIGH
Parameters
Tidal volume:0~1500ml
Respiratory rate:4~99bpm (used in A/C mode)
4~40bpm (used in SIMV mode)
IE ratio:2:1 ~1:4
FiO2:48%~100%
Pressure trigger sensitivity:-2~0kPa (above PEEP)
PEEP (option):0.5~2kPa
Monitoring
Pressure values: Ppeak
Volume/flow values: VT, MV
Time values: Ftotal
Real time curves : Pressure-Time
Alarm
Paw high, Paw low, Apnea, AC power failure, Battery low, O2 deficiencyTechnical data
Screen: LCD
Gas supply: O2, 0.28~0.6MPa
Power supply: AC110~240V, 50Hz/60Hz, <= 65VA
Maximum security pressure: <= 6kPa
Compliance: <= 4mL/100Pa
Noise: <= 65dB(A)
Dimension(HxWxD): 175x170x300mm
Weight : 3.5kgEnvironment requirements
Temperature:
0°C~40°C(Operation)
-10°C~40°C (Storage)
Relative humidity:
≤85%, non-condensing(Operation)
≤80%, non-condensing (Storage)
Atmospheric pressure:
70 ~110 kPa (Operation)
50 ~106 kPa (Storage)
AEonMED - The Operating Expert of OR&ICU