Contact: Chris Zhang(Regional Manager)
Mobile: +86 18588494743(Whatsapp)
Support Alibaba 100% Fully Protection Payment (only After You Get The Parcel, Alibaba Would Release The Funds).
China Supplier Urit Hematology Urit 3000 Plus Hematology Analyzer With Open Reagent System Hematology Cheap Price
Features:
·Large color LCD display & Linux operation system
·Support USB data backup & system upgrade
·Excellent data management with mouse and keyboard
·High efficient self checking system, low maintenance
·Support LIS and HIS with HL7 protocol
·Collaborator of BIO-RADcontrol,target value available
Auto internal & external cleaning for aspiration probe
High precision syringes
68 um and 100 um ruby apertures
Easy to Use
= Large color LCD with high resolution
= 5 shortcut operation keys
= Working status indicator light
= Mouse and keyboard connection for patient data input = Identifiable color reagent connection design
Cost-effective
Tidy fluid system with low reagent consumption # Flexible reagent packaging from original manufacturer = 30+ reagent join venture manufacturers worldwide
Low and Easy Maintenance
= Special self-checking function during initialization
= Auto clog removing by aperture cauterizing and flushing = Error messages
= Openable front panel and side door
Nearly Perfect Software Function
= Friendly interface and illustrated icons
= On board storage up to 100,000
= L-J, X, X-R and X-B QC methods
= Support LIS and HIS connection
= Tailor-made mutli-language operation system
Reliable Spare Parts Quality
= Well-known global spare parts suppliers
= High performance PCB board from fully automatic assembly line
Specifications
·Parameters: WBC, LYM#, MID#, GRAN#, LYM%, MID%, GRAN%, RBC, HGB, HCT, MCV, MCH, MCHC,RDW-CV, RDW-SD, PLT, MPV, PDW, PCT,P-LCR,P-LCC
3-part differentiation of WBC, 21 parameters and 3 histograms
·Principles of operation:WBC/RBC/PLT: Electrical Impedance
HGB: Photoelectric colorimetry
·Aspiration Volume:Whole blood 10uL,Pre-diluted 20uL
·Aperture Diameter:WBC 100um,RBC/PLT 68um
·Throughput:60 samples per hour
·Data storage:More than 100,000 sample results with histograms
·Alarms: Error messages
·Dilution Ration: Whole blood Capillary blood
WBC/HGB 1:232 1:400
RBC/PLT 1:40000 1:45000
·Display: 10.4 inches color LCD
·Control and calibration: L-J,X,X-R,X-B four control modes,manual and auto-Calibration System check
·Dimension: 330mm x 440mm x 500mm
·Power requirement: AC 100V --- 240V, 50/60Hz, 180W
·Input/Output: Support RS232, standard network port and USB
Shortcut keys with standard mouse and keyboard
Internal 57mm thermal printer,
Parallel Port and USB for external printer