Business Type:
Manufacturer/Factory,Trading Company
Business Range:
Pulse Oximeter, Fetal Doppler, Patient Monitor, Sphygmomanometer, ECGOwnershipPrivate OwnerTotal EmployeesAbove 1000 PeopleTotal Annual RevenueconfidentialYear Established1996Certifications(3)ISO9001, ISO14001, ISO13485Product Certifications(2)CE, Test Re
Establishment:
1996
R&D Capacity:
OEM, ODM, Others
Terms of Payment:
LC, T/T, D/P, Paypal, Western Union
Main Markets:
North America, Europe
OEM/ODM Service
Sample Available

Qinhuangdao Contec Medical Systems Co., Ltd.founded in 1992,focuses on researching,manufacturing and exporting of electronic medical equipments. Our manufactory covers an area of 32,000 square, Our pr...

1
    YRS
General Supplier
Supplier Homepage Packaging, Advertising & Office Other Service CONTEC CE MS100 spo2 simulator oximeter simulator spo2 sensor

CONTEC CE MS100 spo2 simulator oximeter simulator spo2 sensor

Get Latest Price
Min. Order / Reference FOB Price
1 Piece US $400.00/ Piece
Local Area: Qinhuangdao, Hebei, China
R&D Capacity: OEM, ODM, Other
Payment Terms: LC, T/T, D/P, Paypal, Western Union
Brand: CONTEC
Place of Origin: Hebei, China
Brand Name: CONTEC
Model Number: MS100

Simulator

Product Description
ConTEC CE MS100 spo2 simulator oximeter simulator
Brief Introduction

MS100 SpO2 Simulator is a kind of Separated SpO2 simulator, small and light. It can perform a series of tests for the oximeter by simulation means, and gives cognizance of veracity about the oximeter.
Because the oximeter manufacturer may use different R-curves, some R-curves used abroad currently have been embedded.
Features

Separated connection between simulator probe and host, operation and test are more convenient.
262K color and 320*240 TFT, adjustable brightness level.
Film key-press, makes the operation more comfortable.
Power with chargeable lithium battery, can display the charge information of battery.
Key volume, can open or close.
Main Functions

Oxygen saturation simulation;
Pulse rate simulation;
Patient state simulation preset;
Can test the reaction time of device;
Can simulate the SpO2 and pulse rate under different amplitude;
Can test the performance of device under different interference source;
Can select different R-curve.
Main Parameters

SpO2
Measuring range: 35%~100%
Resolution:1%
Error: When SpO2 is 75%~100%, the error is the larger between ±2% and ±P(P denotes the precision of the tested oximeter); when SpO2 is 74%~50%, the error is the larger between ±2% and ±P; no define when SpO2 is less than 50%.
Pulse rate
Measuring range:25bpm~250bpm
Resolution:5bpm
Error:1%±1bpm
Amplitude
Measuring range:0.000%~20.000%
Resolution:1% for 1.000%~20.000%; 0.1% for 0.100%~0.900%; 0.025% for 0.000%~0.075%.
Patient state
24 conditions presetting and 8 conditions as default after start. The number of condition can be adjusted by Motion level setup.
Voltage
DC 3.6V~4.2V
Power Supply
Voltage 3.7 rechargeable lithium battery x 1
Battery working life
Charge and discharge no less than 500 times.
Accessaries

A MS100 user manual
A 5.0V adaptor
Certifications
Exhibition
Company Information
BrandCONTEC

Send your message to this supplier

*From:
To:
Contec Medical Systems Co., Ltd.
*Message:

Enter between 10 to 1,000 characters.

*Email code:
This is not what you are looking for? Post a Sourcing Request Now