Features
l15” TFT LCD touch screen displays the ventilation parameters, alarm information and waveforms.
lAir compressor runs stably with low noise, creates quietenvironment for doctorsand patients.
lIntegration breathing circuit design.
lMultiple working modes such as volume control and pressure limit.
lMultiple parameters monitoring interfaceto show clear parameter to the operator, allow operator know the patientconditions in all aspects.
lReal time pressure-time, flow-time loop graphics, pressure volume loop graphicsand high precision ETCO2,O2 concentration detection function included.
lElectrical PEEP control
lBreathing valve isdisassemblyand can be disinfected.
Safety
lThree levelsalarm system, visual and sound alarm information.
lWithmultiple type of alarms, reminding and protection functions.
lSelf-check before running, eliminate system mulfuction.
lAdvanced power management control technology.
lWith built-in backup power supply, whennormalpower supplyshut down, back-up power supplywill start workautomatically.
lSeparatedesign of electric and gas, keep safety running of ventilator.
Specifications
Ventilation modes:
IPPV, A/C, PCV, SIMV, SPONT/CPAP, PSV
Ventilator parameter ranges
Tidal volume(Vt)
0 ~ 2000mL
Frequency(Freq)
Oxygen concentration
1bpm ~ 100 bpm
21%-100%
I/E
4:1~1:8
PEEP
0cmH2O ~ 40 cmH2O
Pressure Limit
20 cmH2O ~ 100 cmH2O
Monitored Parameters
Frequency(Freq)
0 /min ~ 100 /min
Tidal volume(Vt)
0 mL ~ 2500 mL
MV
Airway pressure
Dynamic lung compliance testing
ETCO2 Concentration
0 L/min ~ 99L/min
0 cmH2O ~ 100 cmH2O
1mL/cmH2O ~ 1000mL/cmH2O
0mmHg~152mmHg (0%-20%)
Oxygen concentration
15 % ~ 100 %
Oscillographsdisplay:
P-T(pressure – time)
F-T(flow – time)
P-V loop (pressure – volume loop)
V-T(volume – time )
ETCO2-T (End-tidal CO2 - time)
Working conditions
Gas source
O2,AIR
Pressure
280kPa-600kPa
Voltage
~220V±22V
Power frequency
50Hz± 1Hz
Input power
900VA(with air compressor)
250VA(without air compressor)
Ventilation modes:
IPPV, A/C, PCV, SIMV, SPONT/CPAP, PSV
Ventilator parameter ranges
Tidal volume(Vt)
0 ~ 2000mL
Frequence(Freq)
Oxygen concentration
1bpm ~ 100 bpm
21%-100%
I/E
4:1~1:8
PEEP
0cmH2O ~ 40 cmH2O
Pressure Limit
20 cmH2O ~ 100 cmH2O
Monitored Parameters
Frequence(Freq)
0 /min ~ 100 /min
Tidal volume(Vt)
0 mL ~ 2500 mL
MV
Airway pressure
Dynamic lung compliance testing
ETCO2 Concentration
0 L/min ~ 99L/min
0 cmH2O ~ 100 cmH2O
1mL/cmH2O ~ 1000mL/cmH2O
0mmHg~152mmHg (0%-20%)
Oxygen concentration
15 % ~ 100 %
Graphical display:
P-T(pressure – time)
F-T(flow – time)
P-V loop (pressure – volume loop)
V-T(volume – time )
ETCO2-T (End-tidal CO2 - time)
Alarm and protection
The AC power failure alarm
Power failure or no connection
Internal battery backup low voltage alarm
≤11.3±0.3V
No tidal volume
No tidal volume within 6s
High Minute Volumealarm
LowMinute Volumealarm
5L/min-99L/min
1L/min-30L/min
High Airway pressure alarm
LowAirway pressure alarm
High oxygen concentrationalarm
Low oxygen concentrationalarm
20cmH2O-100cmH2O
0cmH2O-20cmH2O
19%-100%
18%-99%
Continuous pressure alarm
(PEEP+1.5cmH2O) over 16s
Suffocation warning
Fan error
Oxygen deficit
5-60s
Show on screen
Show on screen
The maximum limited pressure
<12.5 kPa
Working conditions
Gas source
O2,AIR
Pressure
280kPa-600kPa
Voltage
~220V ±22V
Power frequency
50Hz± 1Hz
Input power
900VA(with air compressor)
250VA(without air compreesor)