Defibrillator Monitor DM7000
Specification:
Weight: 20 pounds (9.9 kg) (basic configuration).
Size: 13 x 9.6 x 7.2 inches (33 x 24 x 18 cm).
Operating:
Temperature*: 0 to 40° C
Humidity* (NC): 15 to 95% RH
Altitude: Up to 4572 M (15,000 feet)
Transport and Storage:
Temperature: -20 to 55°C
Humidity: 15 to 95% RH (non-condensing)
Atmospheric pressure: 860 to 1060 hPa
Shock/vibration: ISTA 1A
Water: IEC 529, IPxx.
Operating Power: Ni-MH 12 volt - M&B 1.1 Defibrillator Waveform:
Truncated Exponential Biphasic
Output Energy Accuracy:
+/- 15% or 1J (whichever is greater) at 25 to 175 ohms
AED Energy Sequence
150, 200J(twice)
150, 150, 200J(three times)
150,150, 150, 200J(Four times)no battery, 100% AC mains voltage,charge to max energy(360J)
<9 seconds
New battery, ,charge to max energy(360J)
<14 seconds
no battery, 90% AC mains voltage,charge to max energy(360J)
<9 seconds
with a Ni-MH 12 volt - M&B battery after 15 discharges, charge to max energy(360J)
<15 seconds
no battery, 90% AC mains voltage,from system on to charge max energy complete
<20 seconds
with a Ni-MH 12 volt - M&B battery after 15 discharges, from system on to charge max energy complete
<25 secondsno battery, 100% AC mains voltage,charge to max energy(200J)
<7 seconds
New battery, ,charge to max energy(200J)
<9 seconds
no battery, 90% AC mains voltage,charge to max energy(200J)
<7 seconds
with a Ni-MH 12 volt - M&B battery after 15 discharges, charge to max energy(200J)
<10 seconds
no battery, 90% AC mains voltage,from system on to charge max energy complete
<25 seconds
with a Ni-MH 12 volt - M&B battery after 15 discharges, from system on to charge max energy complete
<28 seconds
Analyzing time
9 seconds
The Maximum Voltage
1340±50V
Defibrilator
Resistor
range
AED mode
20Ω to 200Ω
Manual mode
20Ω to 1500Ω
Synchronous delay time
<60ms