Volume
Voltage
Cavity Size mm
Dimension mm
Package Size mm
Optional
LED/LCD
16
220v/50HZ 1700W 616*460*420 USB/Printer
LED/LCD
18
220v/50HZ 1800W
Φ250x360
616*460*420 USB/Printer
LED/LCD
23
220v/50HZ 2000W
Φ250x450
616*460*420
750x545X450
USB/Printer5 - Distiller water outlet
2 - Command panel
6 - Used water reservoir outlet
3 - Power switch
7 - Sterilizer Chamber
4 - USB interface for USB Flash
8 - Safety Door
1.Sterilization level: Class B
2.Sterilization temperature: 121℃,134℃
3.Special sterilization: HIV/AIDS, hepatitis b (HBV), mad cow virus and spore.
4.Drying procedure:A strong negative vacuum drying
5.Operation interface: Adopt the operation of the new interface LCD screen can display temperature and pressure time running state, fault alarm information, etc
6.Test system: Bowie&Dick Steam penetration test, Vacuum test, Helix test
7.Control accuracy: Temperature 0.1℃, pressure 0.1 Kpa.
8.Sterilization data: External USB connection , (printer is optional)
9.Multiple safety: Interlock protection device: Safety valve, Pressure interlock, door alarm, automatic breakdown check. Over temperature and pressure protected.
10.Water system: Internal open water tank, easy to clear.
11.Sterilization Configuration: Activities tray rack, match 3 pallets
AUTOCLAVE DETAILS SHOW:
-Micro-computer control, imported Motorola sensor, high accusation and precision.
-During the sterilizing procedure, the range of temperature is ±0.5.
-Automatic cleaning program in the steamer, avoid the occlusion.
-Display the sterilizing procedure date and curve.
-Water quality monitoring,indicate the water tank condition, automatic notice when used water tank is full.
-Safety valve, double lock, mechanical lock, air lock, multi pressure safe protection device.
-Printer is optional, record the sterilizing procedure with USB.
-Applied with orthopedics, ophthalmology, beauty industries and other area.
MIC Exhibition:
MIC Autoclave (Steam Sterilizer) Applications
1. Beauty
2. Orthopedics
3. Ophthalmology
Attentions:
The process sterilization can consider effective, reliable and repeatable on condition that the material is first suitably treated and subsequently tidy and correctly arranged into the sterilization chamber..
An incorrect arrangement of the load can lead to a difficult and sometimes impossible flowing and/or penetration of the steam on the material,with imaginable consequences. The drying process can strongly be also affected by this factor.
Therefore we suggest some basic directions concerning this aspects, leaving the user to deepen the problem in the most opportune way.