Portable Breathing Apparatus Medical Ventilator, Auto PEEP Machines ventilator Medical ICU Ventilator Machine price
TECHNIQUES SPECIFICATIONSPhysical specifications Display 10.4'' LCD DisplayMode:pneumatically powered and electrically controlled systemTrolley:
Fitted with 4 nos anti-static rubber castors; two of which are lockable for braking and easy
maneuverability with foot operated brake provisionsTechnical Specifications:Gas requirement: Medical oxygen and nitrous oxide with a pressure ranging from 280~600kpasafety valve <12.5 kPaRespiratory Rate1 ~ 99 bpmOxygen Concentration21% ~ 100% Maximum Minute ventilation≥18L / minmodes of ventilationA/C, IPPV, SIPPV, IMV, SIMV, PEET. MANUAL, SIGH, SPONTVentilator compliance≤30 Ml/kPaPEEP:0 ~ 1.0 kPaPtr: -10 ~ 10 hPaIMV frequency:1~20 bpmI/E Ratio:4:1~1:4, Has the inverse ratio ventilationTidal VolumeSetting Range: 50~1500MLDisplay Range: 0~2000ml ; resolution ratio: 10mlInspiratory Plateau:0~1sMaximum Safety Pressure≤ 6.0kPaSigh:
one deep breath per 80 controlled respirations, the inspiration
time is 1.5 times of the setting pointPressure limit range:1.0~ 6.0kPa
Airway pressure alarm: Audible
and visual and with yellow
and red color indicatingLower: 0.1 kPa ~ 6.0 kPa;Upper: 0 ~ 5.0 kPa±0.2 kPaVentilation Volume alarm: ≥ 18L/ MIN Power Supply AlarmAc/dc power supply are after failing to send out alarm immediately
Alarm Time: keep >120s
The airway pressure continues to be higher than 15 hPa ±1 hPa for 15s s, then the machine will raise an audible alarm, the pressure will be displayed in red and the
continuous high pressure red alarm wording is displayed on the screen of the anesthetic respirator.
Air Compressor Power:
AC220V + 22 v, 50 Hz + 1 Hz and 300 va has a good reserve of single-phase ac power supply
FUSE :Φ 5 x 20 / F8A
Output gas pressure: 0.3 ~ 0.4 MPa
Output gas minimum flow: ≥50L/Min Output gas transient flow: ≥100L/Min