Product Description
TECHNIQUES SPECIFICATIONSPhysical specifications Screen: 8.4 '' LCD Display ScreenSuitableAdult & Child Mode:pneumatically powered and electrically controlled systemWorking Mode:Closed; Semi-Closed; Semi-OpenCircuitBreathing circuit integrated standardsFlowmeters:5 Tubes Flowmeters: O2:0.1~10L/Min, N2O:0.1~10L/Min; Air: 0.1~10L / MinTrolley: Fitted with 4 nos anti-static rubber castors; two of which are lockable for braking and easy maneuverability with foot operated brake provisionsDrawer unitTWO drawers totally extractableTechnical Specifications:Gas requirement: Medical oxygen and nitrous oxide with a pressure ranging from O2: 0.32~0.6MPa ;NO2: 0.32 MPa to 0.6 MPa.safety valve ≤12.5 kPaRespiratory Rate1~100bpm
oxygen concentration
in the mixed gas N2O/O2 > 21%Oxygen Flush: 25~75 L/minmodes of ventilationA/C, IPPV, SIPPV, IMV(VCV), SIMV(VCV), PCV, PEEP, MANUAL, SIGHCompliance of the lung≤30 Ml/kPaPEEP:0 ~ 2.0 kPaInspiratory trigger pressure-1.0kPa ~ 2.0 kPaIMV frequency:1~20 times/MinI/E Ratio:8:1 ~ 1:10, Has the inverse ratio ventilationTidal VolumeAdjustable range: 20 ~ 1500 ML, Display range:0~2000MLInspiratory Plateau:0~1sO2 Concentration:21%~100%Waveform display:P----T,F----T,P----V, F---VSigh: one deep breath per 70 ~ 120 controlled respirations, the inspiration time is 1.5 times of the setting pointMaximum safety pressure:≤ 12.5 kPaPressure limit range:0 ~ 6.0 kPa
Airway pressure alarm:
Audible and visual and with
yellow and red color indicatingLower: 0.2kPa~5.0kPa;Upper: 0.3~6.0 kPa±0.2 kPa
Tidal volume alarm: Audible and
visual and with yellow and red
color indicatingupper alarm: 50 to 2000ml, lower alarm: 0~1800ml
Oxygen concentration alarm:
Audible and visual and with yellow
and red color indicatingupper alarm: 21% ~100%; lower alarm:10%~80%Power Supply AlarmAc/dc power supply are after failing to send out alarm immediately
Alarm Time: keep >120s
The airway pressure continues to be higher than 15 hPa ±1 hPa for 15s±1s, then the machine will raise an audible alarm,
the pressure will be displayed in red and the continuous high pressure red alarm wording is displayed on the screen of the
anesthetic respirator.Operating conditionsAmbient temperature: 10 ~ 40oCRelative humidity:no higher than 80 %Atmospheric pressure:860 hPa ~ 1060 hPaPower requirement: 110~220V, 50/60 Hz; Attention: the AC power supply used for the anesthesia machine must be wellgrounded.
Attention: the anesthesia machine used must be equipped with a carbon dioxide monitor complying with ISO 9918:1993,
an oxygen monitor complying with ISO 7767:1997 and an expiratory gas volume monitor complying with 51.101.4.2 of Medical
Electrical Equipment Part II: Special Requirements for Safety and Basic Performance of Anesthesia System. VaporizerVaporizerSetting Range: 0.5%~5%StorageAmbient temperature: -15oC ~ +50oCRelative humidity: no higher than 95 %Atmospheric pressure: 86 kPa ~ 106 kPa. It should be stored in a room without corrosive gas and well ventilatedPackagepackaging boxcomply with the requirement of GB/T 15464
Between the packaging box and product, soft material with appropriate thickness provided to
prevent loosening and mutual friction during transportation Moisture protection and rain protection to ensure that product is protected from natural damage.