Product Description
HB2000 Medical Neonatal Incubator Newborn Baby Care Equipment Transport Infant Warmer
Specifications:
AC power supply: AC220V-230V/50HZ
DC power supply: DC12V/10A or DA24V/6A
Input power: 400VA
Model of control: air mode and baby model controlled by micro-computer
Air temperature control range: 25 ℃-37℃ override mode: 37-38℃
Baby temperature control range: 34-37℃ override model:37-37.5℃
Skin temperature sensor accuracy: ≤0.3℃
Temperature variability: ≤1℃
Temperature uniformity: ≤1.5℃(mattress is on the horizontal position)
Noise inside hood: ≤155dB(A)(environment noise ≤45dB(A))
Failure alarm: over-temperature alarm, temperature deviation alarm, sensor alarm, motor fan failed alarm, power failure alarm and so on
Operating at temperature equilibrium: 90min(set temp 36℃ ambient 15℃)
Environmental temperature
Operating range: 10-30℃(set temp should be higher 3℃ than environmental temp)
Ambient air movement rate: ≤1.0m/s
Package: each unit is packed as two cartons
Size(main body): 109*65.5*76cm
Size(Cabinet): 138*61*40cm
Gross weight(main body): 80kg
Gross weight-Cabinet: 53kg
1. Air mode and Baby mode controlled by micro-computer
2. Use AC Power and DC Power alternatively, DC 12V or DC 24V Power on the ambulance;
3. Set temperature, air temperature, baby temperature and internal battery voltage can be displayed separately;
4. Failure alarm indication;
5. Double wall Acrylic glass hood with side door so that thebassinet can be pulled out from it;
6. Natural air flow humidity;
7. The height of whole unit can be adjusted;
8. With Oxygen cylinder and Oxygen supply configuration;
9. Observation lamp.
Standard configuration:
Main body(including the hood, controller, bassinet, internal battery, and observation lamp) oxygen cylinders, oxygen supply system, skin temperature sensor, IV pole, mattress, adjustable stand.
Optional configuration:>37degree, overrid temperature mode
Product Categories
Company Information
Home