Description:
1.12L, European N class
2. Three Times vacuum
3. Digital display
4. Manual door locking-system
prompt
1. Please place the sterilization on a plane place , which could adjust the height by adjusting foot pad
2. Please adopt distilled water to extend the service life
3. Don't block the heat removal system on the surface of sterilization
4. Please put the instruments on the tray not touching one another allowing space for steam to pass through
5. Please empty the used water tank frequently. Usually it need empty when the clean water in the reservoir is empty.
6. Please ensure the door handle is closed tightly
7. Never try to open the chamber door if and when the pressure display doesn't show "0.0"
8. Please cut off the power and cool down the auto clave before maintenance
9. Don't drag the machine while moving
10. Power connection must be grounded properly
11. Must be provided enough power. (23L:2300VA/18L:1800VA)
12. While the outside temperature is too low, please pre-heating the machine without instruments for 30 minutes
before sterilize
13. Don't cut off the power to stop the machine directly. If you need stop the machine directly, please press the
"START/STOP" button for 3 seconds.
14. The picture in the instruction is for reference only.