JERRY-III Vital Signs Patient Monitor
Introduction
JERRY-III Vital Signs Patient Monitor is a tabletop model and easy to carry. The size of main unit is 280(W)*190(H)*100(D)mm; N.W. is 1kg; The packing size for single model is 33*32*15cm with standard export cartons.This model can be applied in the hospital’s operation room, ICU, clinic section office, Out-patient department, sickroom and emergency treatment.
FeaturesFour parameters: NIBP,SpO2,Pulse Rate,Temperature.SpO2 waveform, PR strength display.2.8 inch color LCD with real time display.Storage/Review trend data up to 3888 groups, data save time interval settable.Audio and visual alarm when parameters exceeded and sensor-off.Built-in rechargeable Li-ion battery, up to 5 hour running time.Automatic shutdown for power saving in 10 mins if no measurement.Via USB to connect with the computer, unique software for real-time monitoring by transmitting data.Battery capacity indicator and prompt when battery low.Bluetooth printer is optional.
Standard accessoriesMain unit 1pcNIBP extension tube 1pcAdult NIBP cuff1pcAdult SPO2 Sensor 1pcSurface temprature probe1pcAC power adaptor 1pcUSB Cable 1pcThe operator’s manual1pc
Optional accessories Pediatric SPO2 sensor Pediatric NIBP cuff Bluetooth Adapter Bluetooth Printer
Specifications
1. NIBPPatient: Adult and PediatricMeasurement method: OscillometricRange:Adult:30mmHg~255mmHg Pediatric: 30mmHg~160mmHgMeasurement unite: mmHg/kPa selectableMeasurement mode: Manual, Auto, STATMeasurement Type: SYS, DIA, MAPMeasurement interval in AUTO mode:1,2,3,5,10,15,20,30,45,60MinMeasuring period in STAT mode: 5 MinResolution:1mmHgOver-pressure protection:Adult:300mmHgPediatric:220mmHg
2. SpO2Patient: Adult, childRange: 0~100%Resolution: 1 %Accuracy: ±2 %(90~99%), ±3 %(70~89%),unspecified (<70%)
3. TEMPChannel: 1Input: Body surface thermal-sensitive resistor temperature sensor Measurement Unit: ℃/F selectableRange: 0~50℃Resolution: 0.1℃Accuracy: ±0.2℃
4. Pulse RateRange: 30~250bpmResolution: 1bpmAccuracy: ±1bpm
5. AlarmSensors off alarm and alarm limit exceeded alarmMode: The parameter’s measurement flashes on the screen, the speaker sounds the alarm and LED flashes on the monitorSetup: User-adjustable high and Low limits
6. Storage andReviewTrend data up to 3888 groups.
7. Display ModeType:240*320 dot matrix color TFT LCD screen and LED screenLCD Display area:44mm*59mmLED Display area: 167mm*79mm
8. Power Requirements
AC power adapterInput: AC 100~240V, 50/60Hz, 0.7AOutput: DC 5V,2.5A
Internal battery Type: 2800mAh/3.7V built-in Li-ion rechargeable batteryOperating Time: 5 hours of stand-by time with a full battery.Recharge time: 8 hours
9. Printer (Optional)Type: Thermal micro-printerPrinter Paper: 58 mm thermal paper,Paper roll diameter≤Φ 33mmInterface: Bluetooth wireless interfaceRecording Types: Real-time printing and review data printingPower Supply:
Rechargeable battery 1700mAh~2000mAh.7.2V.
Charge(discharge) ≥1000times.
Charge time: 2hoursWork Operation/relative humidity:0℃~55℃/10%~80%RHStorage Operation/relative humidity:–20℃~60℃/10%~90%RH