JR2000 Multi-para patient monitor
Outlook
JR2000 patient monitor is easy to carry, with 800*600, 12.1inch color TFT-LCD screen, size of main unit is 280*350mm, thickness is 140mm, weight of main unit with battery is 3.5KG. Packing carton of each unit is 44*29*39CM,weight of patient monitor with accessories and packing material is 7KG.
Featuresbattery working time is at least 4hours, recharge time is 2.5hours. Suitable for medical field rescue, ambulance, move and transit.display mode of big font, trend review, ECG 7 lead, support one key operation, touchable screen is optional. Suitable for all working places, can used for neonate, child, adult and veterinary; support mode of diagnose,monitor and operation.15 type arrhythmic analysis,7-lead ECG waveform display in phase. 15 type drug calculation and titration table.efficient resistance to interference of defibrillator and electrosurgical cautery, respiration can choose III,meet requirement of operation, ward.Use digital SpO2 technology,can test for 0.1% weak,high accuracy.anti-ESU,anti-defibrillator,RA-LL impedance respiration,digital filtering.Double over-pressure NIBP protect,temperature drift.Optional functions:Touchable screen, EtCO2, 2-IBP, Printer, Central Monitoring System
Standard accessoriesPatient Monitor main unit(include battery)1pc5-lead ECG cable 1pcsurface temp.probe 1pcadult NIBP cuff 1pcNIBP tube1pcadult SpO2 sensor 1pcpower cord1pcelectrode 5pcsUser manual 1pc
Optional accessoriesLi-ion batteryadult,pediatric,neonate(fingertip,finger stall,wrap) SpO2 sensoradult,pediatric,neonate NIBP cuffvarious CO2 sampling pipesDisposable IBP transducer and reusable IBP cable3-lead,5-lead ECG cable,clip ECG cable in Euro or USA standardsurface and rectal temp.probe
Specifications
1.ECGLead Mode: 5 Leads (R, L, F, N, C or RA, LA, LL, RL,V)Lead selection: I, II, III, avR, avL, avF, V, Waveform: 2 chLead mode: 3 Leads ( R, L, F or RA, LA, LL)Lead selection: I, II, III, Waveform: 1 chGain: ´2.5mm/mV, ´5.0mm/mV, ´10mm/mV, ´20mm/mV, auto
HR and AlarmRange
Adult: 15 ~ 300 bpm
Neo/Ped: 15 ~ 350 bpmAccuracy: ± 1% or ± 1bpm,which greatResolution: 1 bpmSensitivity: > 200 (uV P-P) Differential Input Impedance: > 5 MΩCMRR
Monitor: > 105 dB
Operation: > 105 dB
Diagnosis: > 85 dBElectrode offset potential: ±300mVLeakage Current: < 10 uAbaseline Recovery: < 3 S After Defi.ECG Signal Range: ±8 m V ( Vp-p )Bandwidth
Surgery: 1 ~ 15 Hz
Monitor: 0.5 ~ 35 Hz
Diagnostic: 0.05 ~ 100 HzCalibration Signal: 1 (mV p-p), Accuracy:±5% ST Segment Monitoring Range: Measure and Alarm -2.0 ~ +2.0 mVARR DetectingType: ASYSTOLE, VFIB/VTAC, COUPLET, BIGEMINY, TRIGEMINY, R ON T, VT>2, PVC, TACHY, BRADY, MISSED BEATS, PNP, PNCAlarm: AvailableReview: Available
2.RespirationMethod: Impedance between R-F(RA-LL)Differential Input Impedance: >2.5 MΩMeasuring Impedance Range:0.3~5.0Ωbase line Impedance Range: 0 ~ 2.5 KΩBandwidth: 0.3 ~ 2.5 Hz
Resp. RateMeasuring and Alarm Range
Adult: 0 ~ 120 rpm
Neo/Ped: 0 ~ 150 rpmResolution: 1 rpmAccuracy: ±2 rpmApean Alarm: 10 ~ 40 S
3.NIBPMethod: OscillometricMode: Manual, Auto, STATMeasuring Interval in AUTO Mode: 1, 2, 3, 4, 5, 10, 15, 30, 60, 90, 120, 180, 240,480 (Min)Measuring Period in STAT Mode: 5 MinPulse Rate Range: 40 ~ 240 bpmAlarm Type: SYS, DIA, MEANMeasuring and alarm range
Adult Mode
SYS: 40 ~ 270 mmHg
DIA: 10 ~ 215 mmHg
MEAN: 20 ~ 235 mmHg
Pediatric Mode
SYS: 40 ~ 200 mmHg
DIA: 10 ~ 150 mmHg
MEAN:20 ~ 165 mmHg
Neonatal Mode
SYS: 40 ~ 135 mmHg
DIA:10 ~ 100 mmHg
MEAN:20 ~ 110 mmHgResolution Pressure:1mmHgAccuracy Pressure Maximum Mean error: ±5mmHg
Maximum Standard deviation: ±8mmHgOverpressure Protection
Adult Mode: 297±3 mmHg
Pediatric Mode: 240±3 mmHg
Neonatal Mode: 147±3 mmHg
4.SpO2Measuring Range: 0 ~ 100 %Alarm Range: 0 ~ 100 %Resolution: 1 %Accuracy: 70% ~ 100% ±2 %
0% ~ 69% unspecifiedActualization interval: about 1 Sec.Alarm Delay: 10 Sec.
Pulse Rate Measuring and Alarm Range: 20~300bpmResolution: 1bpmAccuracy: ±2bpm
5.TEMPChannel: 1Measuring and Alarm Range: 0 ~ 50 ℃Resolution: 0.1℃Accuracy: ±0.1℃Actualization interval: about 1 Sec.Average Time Constant:<10 Sec.
6.Power Supply
AC Power
100~240V,50/60Hz,10VA