YSAV400A
Medical ICU Ventilator With Air Compressor TECHNIQUES SPECIFICATIONSPhysical specifications DisplayLED DisplayMode:pneumatically powered and electrically controlled systemTrolley: Fitted with 4 nos anti-static rubber castors; two of which are lockable for braking and easy maneuverability with foot operated brake provisionsTechnical Specifications:Gas requirement: Medical oxygen with a pressure ranging from 280~600kpasafety valve <12.5 kPaRespiratory Rate6~60 times/ minOxygen concentration output<50%minute volume≥18L modes of ventilationA/C, IPPV, SIPPV, IMV, SIMV, PEET. MANUAL, SIGHPEEP:0.1 ~ 1.0 kPaPtr: -4 ~ 10 hPaIMV frequency:1~12 bpmI/E Ratio:1:1.5; 1:2.0; 1:2.5; 1:3.0Tidal VolumeNot less than 50~ 1200mlInspiratory Plateau:0~1sMaximum Safety Pressure≤ 6.0kPaSigh: one deep breath per 80 controlled respirations, the inspiration time is 1.5 times of the setting pointPressure limit range:0 ~ 6.0 kPaAirway pressure alarm: Audible and visual and with yellow and red color indicatingLower: 0.9 kPa ~ 5.4 kPa;Upper: 0.5 kPa±0.2 kPaVentilation Volume alarm: ≥ 18L/ MIN Power Supply AlarmAc/dc power supply are after failing to send out alarm immediately
Alarm Time: keep >120sThe airway pressure continues to be higher than 15 hPa ±1 hPa for 15s±1s, then the machine will raise an audible alarm, the pressure will be displayed in red and the continuous high pressure red alarm wording is displayed on the screen of the anesthetic respirator.Operating conditionsAmbient temperature: 10 ~ 40oCRelative humidity:no higher than 80 %Atmospheric pressure:860 hPa ~ 1060 hPaPower requirement: 100-120 Vac, 50/60 Hz; Attention: the AC power supply used for the ICU ventilator must be wellgrounded.StorageAmbient temperature: -15oC ~ +50oCRelative humidity: no higher than 95 %Atmospheric pressure: 86 kPa ~ 106 kPa. It should be stored in a room without corrosive gas and well ventilatedPackagepackaging boxcomply with the requirement of GB/T 15464 Between the packaging box and product, soft material with appropriate thickness provided to prevent loosening and mutual friction during transportation Moisture protection and rain protection to ensure that product is protected from natural damage. Safty & AlarmGas source pressureImmediate alarm; “high level alarm” indicator flickers raising an audible alarm.Airway Pressure AlarmThe airway pressure exceeds 15hpa. The maximum delay is no longer than 17s. The “high level alarm” indicator flickers raising an audible alarm. Tidal Volume AlarmLower than 50ml, The alarm will be raised after one breathing cycle. Ventilation Volume AlarmLower: 0~12L/Min;Up: 18L/ MinBattery Voltage AlarmIt alrms when the battery voltage is lower than 10.5V.Power On Self TestSelf test before working, assure it's working normally. STANDARD CONFIGURATIONSQTYNAME1 setMain unit1 setOxygen tube2 setSilicone Mask1 pcPower line2 pcsThreaded pipe2 pcsStandard Pipe1 setOxygen Reducer1 setHumidifier1 pcBattery1 setflow sensor2 pcsA leather bag ( Blue)1 setTools with the machine1 setUser Manual (English Version)medical ventilator