This model is special for training usage in medical college, hospital and etc.it can't be used on human.
The ventilator is aerodynamic, electronic-controlled ventilator, functions can be set through the respiratory rate, inspiratory time, tidal volume parameter to achieve management of the patient breathe, There are Japanese Sharp large screen true color LCD to show waveform and parameters, Germanic Baode proportion electromagnetism valve is suction control valve, Japanese SMC electronic air and oxygen mixer and etc. These high quality imported parts bring out the machine is more accurate, more reliable and more stable.
Main technical parameters and performance
1. Respiratory Mode: A/C, SIMV, PSV,SPONT, MMV, IPPV, SIGH, ,PEEP
2. Parameter setting up:
2.1 Tidal volume (Vt): 0 ~ 2000ml
2.2 Breathing frequency (f): 4 to 99 times / minute
2.3 SIMV Frequency: 6 ~ 60 times / minute
2.4 Inspiratory time (Ti): 0.2 ~ 6 s
2.5 Breath-hold time (Tp): 0 to 2 s
2.6 Trigger pressure (Pt): -2kPa~ 2kPa,
2.7 PEEP pressure: 0.1 ~ 2kPa,
2.8 PSV pressure: 0~ 5 kPa
2.9 Safety largest gas pressure: ≤6 kPa
2.10 Breathing oxygen concentration: 40 ~ 100%, consecutive adjustable
4. Overall performance:
4.1 Drive mode: aerodynamic, electronic-controlled
4.2 10.4 inch colorful TFT screen
4.3 Power supply: AC 100V--230V.
4.4 Aerodynamic source pressure: 0.28~0.6MPa
4.5 mainframe input power: 65VA.
4.6 complete machine noise: ≤65 dB (A).
4.7 Compliance system: ≤4 ml/100Pa