200w dual output 60v 12v power supply for ventilators infusion pumps medical equipment
Product Description
Features:
1, Wide range AC input voltage;
2, PFC Function;
3, High efficiency,Low interference;4, Short circuit, overload, overvoltage, overheating protection.
parameter Item No. K18N-UP200D60+12CH1 CH2
OUTPUT
DC Output Voltage60V12V
Rated Current3A2A
Current Range0.2-3A0.1-2A
Rated power180W24W
Ripple & Noise600mVp-p150mVp-p
DC volltage ADJCH1:±10%
Output v.Tolerance±1%±8%
Line Regulation
±0.5%,±2%
Load Regulation
±0.5%,±6%
Setup,Rise,Hold up time
500ms,50ms,20ms
INPUT
Input Voltage range100-240V AC
Input current
1.2A/230Vac 2.4A@115Vac
Input frequency
47-63Hz
Efficiency>88% 220V ACInrush currentCold start 20A/230VacLeakage current<0.35mA 240V ACPower factorPF>0.98/115V & PF>0.95/230V
Over Temp.protectionOver load protection105~150%,Hiccup,automatic recoveryShort circuit protectionShut down,automatic recoveryOver voltage CH1: 64-70VOver temp. protectionThe cover temp.>105℃,output shut down and automatic recovery
ENVIRONMENT
Working Temp.,Humidity
-20~50℃,10~90% RH,No condensation
Storage Temp.,Humidity
-20~85℃,10~95% RH
Temp.Coefficeent
±0.03%/℃(0~50℃)
Vibration
10~500Hz, 2G 10min./1cycle, PERIOD FOR 60min. EACH AXES
SAFETY&EMC
Safety standardsGB4943,EN60950Withstand Voltage
I/P-O/P:1.5KVAC I/P-FG:1.5KVAC O/P-FG:0.5KVAC
Insulation resistanceI/P-O/P, I/P-FG, O/P-FG:500VDC / 100M Ohms
EMC standardsEN55022 CLASSB,EN61000-3-2
OTHER
Dimension
127*82*45mm(L*W*H)
Weight
0.4Kg
1. All parameters 220VAC input, rated load, ambient temperature 25 ℃, measured under conditions of 70% humidity;
2. Accuracy: includes set up tolerance, line regulation, load regulation and so on;
3. Ripple Test: 30cm long strands connect the power supply and the load, the load side and 0.1u and 47u capacitor to
20MHz oscilloscope measured at the load end;
4. Line regulation: at rated load conditions, the input voltage changes from low to high measured;
5. Load regulation: the measured output load from 0 to 100% load change measured, other load is fixed at 60% load.
Related Products
Company Information
Packaging & Shipping
Payment&Shipping
Payment
We accept T/T, L/C,or Western Union,if you have any
problems about payment,please contact us
Shipping
15-20 days after sample approval and deposit received.
Tracking No. will be sent to you after delivery and you
will get every important shipping status by email
FAQ
Q1. Can I have a sample order for power supply?
A1: Yes, we welcome sample order to test and check quality. Mixed samples are acceptable.
Q2: Do you accept OEM/ODM for the switching power supply ?
A2: Yes. we do.
Q3:How about the quality?
A3:We have the better professional and experienced engineers team and strict QA and QC system.Repair rate under 0.1%~0.2%
Q4:What is PFC?
A4:PFC stands for Power Factor Correction. The purpose of PFC is to improve the ratio of apparent power and real power. The power factor is only 0.5~0.7 in non-PFC models. In PFC models, the power factor can reach above 0.9. The calculation formulas are as below:
Apparent Power=Input Voltage x Input Current (VA)
Real Power= Input Voltage x Input Current x Power Factor (W)